Carbide Burs Cleaning Instructions

CLEANING

STEP 1. Wear gloves when handling contaminated instruments. Pre-soak carbide burs in a container of soapy water to loosen debris. Ultrasonic systems may also be used to loosen debris in burs, however burs should be separated from each other in a bur block during immersion to prevent damage.

STEP 2. Brush away remaining debris using a SS White stainless steel wire brush (order no. 26040) and rinse burs under running water.

STEP 3. After rinsing, dry burs thoroughly by placing them on absorbent towels. Pat dry all surfaces.

1. Dry Heat Sterilizers* 170°C (340°F) for 1 hour.

*These methods, when used according to manufacturers instructions,will not corrode or dull carbide burs.

2. Steam autoclaves 121°C (250°F) for 20 minutes @ 15 p.s.i. These systems will effectively sterilize carbide burs, however, potential for corrosion is present.

STERILIZATION

Proper sterilization of carbide burs is extremely important because it eliminates the threat of cross infection of patients and staff with communicable disease organisms. AVOID cold sterilizing solutions as they contain oxidizing agents which may weaken carbide burs. AVOID worn chucks which may cause bur slippage and/or vibration that can lead to breakage.